Gas Calibration Standards
Which Calibration Gas Standard is Right for Your Application?
Parameter | Wafer | Dynacal Perm Tubes | G-Cal | Bottled Gas |
---|---|---|---|---|
Lifespan | 12 months from date of shipment | 12 months from date of shipment | 12 months from date of shipment | Depending on the chemical, 1 year from fill date in most cases |
Handling | No permitting or dangerous goods handling required | No permitting or dangerous goods handling required | No permitting or dangerous goods handling required | May require permitting, specialized dangerous goods handling procedures |
Environment | Laboratory | Laboratory | Field / Laboratory | Laboratory |
Temperature Requirements | Require temperature-controlled environment 30° – 50°C | Require temperature-controlled environment 30° – 110°C | Can be operated at ambient temperature. No oven or Dynacalibrator required. Wafer devices available at 50°C | Not temperature sensitive - can be operated at room temperature. |
Size | Length 5.5 cm, Diameter 1.6 cm | Small for flexibility and convenience. Maximum length of 23.5cm. Maximum diameter 1 cm | 1.0" W x 4.25" L, 1.0" W x 6.25" L, 1.0" W x 8.0" L (Gas Phase) Depending on Chemical and Rate. | Bulky, takes up valuable lab space |
Arsine and Phosphine compatible | No | No | Yes. Limited Rates (200 ng/min – 1,000 ng/min) | Yes - But very difficult to obtain. Extremely expensive. DOT/ICAO Security Plan may be Required. |
Chemicals | H2S SO2 , COS, NO2, NH3 | More than 500 available, including experimental by request | More than 100 available, including developmental chemicals by request | Mutually reactive chemicals cannot be used |
Customizable mixtures | Mixtures can be created by using (chemical compatible) individual devices in a chamber. Vici provides pure chemicals only; no mixtures. | Mixtures can be created by using (chemical compatible) individual devices in a chamber. Vici provides pure chemicals only; no mixtures. | Mixtures can be created by using (chemical compatible) individual devices in a series. Vici provides pure chemicals only; no mixtures. | Expensive custom mixtures must be purchased, or keeping a range of expensive gas cylinders in stock. |
Permeation rates | Ideal for high vapor pressure substances and low permeation rates, maximum 2500 ng/min | Variable, typically 30 ng/min to 4,000 ng/min | < 200 ng/min to 30,000 ng/min depending on temperature | Set and unchangeable. |
NIST traceable | Yes | Yes | Yes | Unknown traceability standards |
Cost | Economical, low cost to ship and store. With Dynacalibrator, break even and cost-savings delivered by Year 2 | Economical, low cost to ship and store. With Dynacalibrator, break even and cost-savings delivered by Year 2 | Economical, low cost to ship, no Dynacalibrator or oven required | High costs, including cylinder purchase, rental, shipment, disposal |
Stability / reliability | Stable and consistent. | Stable and consistent. | Stable and consistent. | Standard can degrade within the cylinder due to changes in concentration, pressure, and interaction of calibration surfaces and components |
Concentrations | Customizable, based on dilution flow rate and/or set point temperature | Customizable, based on dilution flow rate and/or set point temperature | Customizable, based on dilution flow rate and/or set point temperature | Fixed non-changeable |
Accuracy | 98% Accuracy | 98% Accuracy | 95% Accuracy | Unknown accuracy |